Agilent InfinityLab LC Series Vialsamplers User Manual 106
3Using the Module
Using the Optional Integrated Column Compartment
3 Click the Insert button to insert a line above the currently selected line, or the
Append button to add a line to the end of the table.
The new line contains a copy of the information in the template line.
4 Click the Edit button to display a dialog box that allows you to edit the
column-specific information such as Serial Number, Batch Number and
Description.
5 Add or edit the other column-specific information (for example, maximum
pressure, maximum temperature, length, diameter, particle size) in the Edit
Columns table.
6 If the column is installed and will be used in the Multicolumn Thermostat,
select YES in the Installed column.
7 Click the Plumbing button.
The Column Assignment dialog box is displayed.
Figure 21 Column Assignment dialog box