Agilent InfinityLab LC Series Vialsamplers User Manual 267
8Maintenance
Exchange the Wash Port Assembly
17 Align the needle tip with the seat and ensure that they
are concentric. If needed, carefully bend the needle into
the right position with your fingers.
18 Reconnect the sample loop to the needle. Make sure of
the correct positioning of the loop capillary, the
uncoated part of the capillary must be horizontal.
NOTE
Do not overtighten the fitting!
NOTE
Incorrect positioning and installation of the needle/loop
connection can result in damaging or breaking the sample
loop.
19 In Lab Advisor use Next to lift the needle slowly into the
up position.
OR
In the Local Controller, move the needle up to the upper-
most position.
20 Move the wash port back into its normal position by
turning it to the right.

Table of Contents

Other manuals for Agilent Technologies InfinityLab LC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfinityLab LC Series and is the answer not in the manual?

Agilent Technologies InfinityLab LC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinityLab LC Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals