EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>InfinityLab LC Series

Agilent Technologies InfinityLab LC Series User Manual

Agilent Technologies InfinityLab LC Series
369 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #268 background imageLoading...
Page #268 background image
Agilent InfinityLab LC Series Vialsamplers User Manual 268
8Maintenance
Exchange the Wash Port Assembly
21 Install the safety cover (1.) and fix the screw (2.). 22 In Lab Advisor, select Next to move the needle into the
needle seat and then Back to finish the maintenance
procedure.
OR
In the Local Controller, select Continue to move the nee-
dle into the needle seat and then Exit to finish the main-
tenance procedure.
23 Select Back to finish the procedure.
24 Close the doors.
Next Steps:
25 Purge the wash port. Check if solvent is delivered from
the peristaltic pump.
26 Perform a System Pressure Test.
(1.)
(2.)

Table of Contents

Other manuals for Agilent Technologies InfinityLab LC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfinityLab LC Series and is the answer not in the manual?

Agilent Technologies InfinityLab LC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinityLab LC Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals