Agilent InfinityLab LC Series Vialsamplers User Manual 83
3Using the Module
Install the Optional Integrated Column Compartment
3 Connect the injector valve (port 6) of the Vialsampler to
the inlet port of the ICC heat exchanger with a suitable
capillary (see
“Capillary Connections” on page 278 for
recommendations).
4 Connect a suitable capillary to the respective outlet port
of the ICC heat exchanger (see
“Capillary
Connections” on page 278 for recommendations).
5 Tighten the capillary fitting to make the connection
pressure tight.
NOTE
Do not overtighten the fitting (maximum torque 2.5 Nm).
NOTE
The use of capillaries with SL or SX fittings are recom-
mended for a better handling.
6 Mount the column holders on the heat exchanger.
max.
2.5 Nm
max.
2.5 Nm