EasyManua.ls Logo

Agilent Technologies InfinityLab LC Series - Page 84

Agilent Technologies InfinityLab LC Series
369 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent InfinityLab LC Series Vialsamplers User Manual 84
3Using the Module
Install the Optional Integrated Column Compartment
7 Connect the column to the free end of the capillary
coming from the ICC heat exchanger. For best perfor
-
mance, use capillaries with the InfinityLab Quick Con-
nect Fitting.
a Turn the column onto the Quick Connect Fitting
until the first resistance is observed.
b Close the lever to establish a pressure tight connec-
tion.
8 Connect a suitable capillary to the outlet port of the col-
umn. For best performance, use capillaries with the
InfinityLab Quick Turn Fitting.
a Connect the capillary with the Quick Connect Fitting
and tighten it until finger-tightness.
The fitting should be pressure tight up to 600 bar.
b If needed, tighten the fitting further with the help of
two suitable wrenches.

Table of Contents

Other manuals for Agilent Technologies InfinityLab LC Series

Related product manuals