166 Chapter 3
Mode
Two Port Insertion Loss Measurement
Mode
Avg Number
Sets the number of measurement averages used when calculating the measurement result.
Clear Normalize
Clear Normalize deletes the current normalization data.
Limits
Access the limit mask keys. You can define a limit mask that matches your maximum or
minimum measurement requirements. This mask can then be compared with the
measured trace data to determine pass/fail conditions.
Example TPIL:AVER:TCON RMAX
Key Path
Meas setup, Avg Mode, Repeat Max Hold
Mode Stimulus / Response
Remote Command [:SENSe]:TPILoss:AVERage:COUNt <integer>
[:SENSe]:TPILoss:AVERage:COUNt?
Example TPIL:AVER:COUN 50
Preset 4
Min 0
Max 100
Key Path
Meas Setup, Avg Number
Mode Stimulus / Response
Remote Command
Notes
Front Panel feature only.
There is no remote command to clear the normalization.
Working remotely, you will use the current normalization data.
Or else you must perform another normalization if you do not
want to use the current one.
Key Path
Meas Setup, Clear Normalize