Chapter 3 197
Mode
One Port Insertion Loss Measurement
Mode
Level: 2nd Segment Left
Sets the power level, for the 2
nd
segment to the Left of the Center segment. This key is only available when 2
nd
has been selected with the Segments key.
Dependencies /
Couplings
The span for each of the left
The Left span is calculated as follows:
(<maximum span>/2) – (<center segment span>/2) - <first
segment left span>
The Right span is calculated as follows:
(<maximum span>/2) – (<center segment span>/2) - <first
segment right span>
Where:
<maximum span> is the maximum frequency span of the
instrument (3 GHz or 6 GHz)
<center segment span>
<first segment left span>
<first segment right span>
Preset Not affected by a preset but set per default to Left after a power
cycling
State Saved Saved in instrument state.
Range Left | Right
Key Path
Meas Setup, Limits, Limit Masks, Limit Mask 1, 2nd Segment (Left or
Right Limits)
Mode Stimulus / Response
Notes
Key is only displayed when the selected segment is 2
nd
, and the
selected segment side is Left.
Preset Not affected by a preset but set per default to 50 dB after a
power cycling
State Saved Saved in instrument state.
Min –100 dB
Max 100 dB
Key Path
Meas Setup, Limits, Limit Masks, Limit Mask 1, 2nd Segment: Left
Level