Chapter 3 219
Mode
Return Loss
Mode
Select Trace 
Selects which trace the menu keys will affect. Press 
Select Trace until the number of the 
desired trace is underlined.
Update On/Off (View)
This key lets you turn 
Off active trace updating. That is, make it an inactive trace (or a 
stored trace). This does not affect whether the trace is visible or not. Use the 
Display 
Show/Blank key to change its visibility.
• When a trace becomes inactive, the following things happen:
• Updates to the trace stop after the current sweep is complete. 
• The trace is displayed at half intensity (a long as it stays inactive)
• Inactive traces display across the entire x axis of the instrument. Their horizontal 
placement does not change even if x axis settings subsequently are changed, although 
y-axis settings will affect the vertical placement of data. 
• In most cases, inactive traces are static and unchanging; however, there are cases when 
an inactive trace will update, specifically:
— If data is written to that trace from remote
— If trace data is loaded from mass storage
When a trace becomes active (
Update = On), the trace is cleared, the average count is reset, 
and the instrument prepares to begin a new sweep. 
Preset Trace 1
State Saved The number of the selected trace is saved in Instrument State 
Range 1 | 2
Key Path
Trace/Detector, Select Trace 
Dependencies / Couplings Setting the trace type (even to the type it was already in) puts the 
trace in 
Update = On
Preset On for Trace 1; Off for 2
State Saved Saved in Instrument State 
Key Path
Trace/Detector, Update On/Off (View)