Chapter 3 237
Mode
Distance To Fault
Mode
Level: Center
Sets the power level for the center segment of the current selected limit mask. This key is available
only when Center has been selected with the Segments key.
Span: Center
Sets the frequency span for the center segment of the currently selected limit mask. This key is only
available when Center has been selected with the Segment key.
Dependencies /
Couplings
If Off no limit mask will be displayed even if the 1
st
or 2
nd
segments are turned On.
If On the center segment is drawn on the display centered
round the current center frequency.
Preset OFF
State Saved Saved in instrument state.
Range On | Off
Key Path
Meas Setup, Limit Masks, Limit Mask 1, Center Segment
Mode Stimulus / Response
Restrictions and Notes Key is only displayed when the selected segment is Center.
Preset Not affected by a preset but set per default to 40 dB after a
power cycling
State Saved Saved in instrument state.
Min –10 dB
Max 150 dB
Key Path
Meas Setup, Limit Masks, Limit Mask 1, Center Segment Level
Default Unit dB
Mode Stimulus / Response
Restrictions and Notes Key is only displayed when the selected segment is Center.
Preset Not affected by a preset but set per default to 2 MHz after a
power cycling
State Saved Saved in instrument state.