238 Chapter 3
Mode
Distance To Fault
Mode
1st Segment
Sets the state of the first segment to On or Off. This key is available only when
1st has been selected
with the
Segments key.
1st Segment (Left or Right Limits)
Min 0 m (if Unit is Meters)
0 ft. (if Unit is Feet)
Max 304.8 m (if Unit is Meters)
1000 ft. (if Unit is Feet)
Key Path
Meas Setup, Limit Masks, Limit Mask 1, Center Segment Span
Default Unit m (meters) if the “Unit” parameter is set to “Meter”
ft. (feet) if the “Unit” parameter is set to “Feet”
Mode Stimulus / Response
Notes
Key is only displayed when the selected segment is 1
st
.
Dependencies /
Couplings
If Off the 2
nd
limit mask segment will not be displayed even if
turned On.
If On and the Center segment is On:
The left hand side 1
st
segment is drawn to the left hand side of
the center segment at the specified power level
The right hand side 1
st
segment is drawn to the right hand side
of the center segment at the specified power
For both left and right 1
st
segments when On
If the 2
nd
segment is Off the 1
st
segment extends so the total
limit mask span is equal to the current frequency span.
If the 2
nd
segment is On the left and right 1
st
segments use the
specified span values.
Preset Off
State Saved Saved in instrument state.
Range On | Off
Key Path
Meas Setup, Limit Masks, Limit Mask 1, 1st Segment