Chapter 3 239
Mode
Distance To Fault
Mode
Sets 1st Segment to allow definition of the limits Left or Right of the center segment. The Left or
Right amplitude limits can be changed with the
Level key. The Left or Right span limits can be
changed with the limits menu
Span key. This key is only available when 1st has been selected with
the
Segments key.
Level: 1st Segment Left
Sets the power level, for the 1
st
segment to the Left of the Center segment. This key is only
available when
1st has been selected with the Segments key.
Level: 1st Segment Right
Sets the power level, for the 1
st
segment to the Right of the Center segment. This key is only
available when
1st has been selected with the Segments key.
Mode Stimulus / Response
Notes
Key is only displayed when the selected segment is 1
st
.
Preset Left [segment]
State Saved Saved in instrument state.
Range Left | Right
Key Path
Meas Setup, Limit Masks, Limit Mask 1, 1st Segment (Left or Right
Limits)
Mode Stimulus / Response
Notes
Key is only displayed when the selected segment is 1
st
, and the
selected segment side is Left.
Preset Not affected by a preset
State Saved Saved in instrument state.
Min –10 dB
Max 150 dB
Key Path
Meas Setup, Limit Masks, Limit Mask 1, 1st Segment: Left Level
Default Unit dB
Mode Stimulus / Response