Chapter 3 241
Mode
Distance To Fault
Mode
2nd Segment
Sets the state of the first segment to On or Off. This key is available only when 2
nd
has been
selected with the Segments key.
Notes
Key is only displayed when the selected segment is 2
nd
, and the
selected segment side is Right.
Dependencies /
Couplings
If the Center, 1
st
, and 2
nd
segments On, the right 2
nd
segment
span adjusts on a change in the right 1
st
segment span.
Preset Not affected by a preset but set per default to 20 m (meters)
after a power cycling.
State Saved Saved in instrument state.
Min 0 m (if Unit is Meters)
0 ft. (if Unit is Feet)
Max 304.8 m (if Unit is Meters)
1000 ft. (if Unit is Feet)
Key Path
Meas Setup, Limit Masks, Limit Mask 1, 1st Segment: Left Span
Default Unit m (meters) if the “Unit” parameter is set to “Meter”
ft. (feet) if the “Unit” parameter is set to “Feet”
Mode Stimulus / Response
Notes
Key is only displayed when the selected segment is 2
nd
.
Dependencies /
Couplings
If Off the 2
nd
limit mask segment will not be displayed even if
turned On.
If On, and the First and Center segments are On:
The left hand side 2
nd
segment is drawn to the left hand side of
the left first segment at the specified power level
The right hand side 1
st
segment is drawn to the right hand side
of the right first segment at the specified power
The segment spans are calculated as documented in section
above
Preset Off
State Saved Saved in instrument state.
Range On | Off