Chapter 3 89
Mode
Meas Off (Spectrum Analyzer) Measurement
Mode
Frame Skip Enables you to set the number of frames you would like to skip when capturing
data. You may set this value from 0 to 2,147,483,647 frames. Increasing the frame skip
value causes the display to redraw the spectrum every nth frame and a block of lines are
shown at once instead of a single line at a time. Higher frame skip values are for use with
fast measurements.
Palette Switches the spectrogram color mode between full color and grayscale.
Display Enable (Remote Command Only)
Turns the display on/off, including the display drive circuitry and the backlight. The three
goals of turning the display off are:
To increase speed as much as possible by freeing the instrument from having to update
the display
To reduce emissions from the display, drive circuitry and backlight
Preset Max Speed (0 s.)
State Saved Saved in State
Min 400 nS
Max (200*1024)/(2*RBW)
Key Path
View/Display, Spectrogram, Update Interval
Mode SA
Preset 0 frames
State Saved Saved in State
Min 0
Max 2,147,483,647
Key Path
View/Display, Spectrogram, Frame Skip
Mode SA
Preset Full Color
Range Full Color | Grayscale
Key Path
View/Display, Spectrogram, Palette