MM-400A
8. Operation Screens
8-11
FORCE PEAK
*1
Shows the peak force including the outside of the measurement range.
FORCE AVG1
*1
You can specify two measurement ranges for a single force application.
Shows the mean force over the force measurement interval 1. (START TIME
1 and END TIME 1 in the EXTEND SETUP (4) screen)
FORCE AVG2
*1
You can specify two measurement ranges for a single force application.
Shows the mean force over the force measurement interval 2. (START TIME
2 and END TIME 2 in the EXTEND SETUP (4) screen)
FORCE INITIAL
*1
Measures and shows the force before welding. Select CURRENT START or
EXTERNAL INPUT for the timing setting INITIAL MEASUREMENT in the
EXTEND SETUP (5) screen.
FORCE FINAL
*1
Measures and shows the force after the end of current flow (between the
final current flow and the end of final delay time).
FORCE REAL TIME
*1
The MM-400A constantly measures and displays the force while MEAS is
selected. The MM-400A stops measurement when PROG is selected.
Measurement is made at intervals of twice a second. If you wish to select
FORCE REAL TIME for measurement, select REAL TIME for TRIGGER the
BASIC SETUP (1) screen.
FORCE TIME
*1
Shows the time from when the force signal exceeds TRIGGER LEVEL to
when the signal falls below TRIGGER LEVEL.
EXT PEAK
*1
Shows the peak external input voltage or current (±10 V voltage or 4 to 20
mA current) converted at the preset conversion factor.
EXT AVG1
*1
Shows the mean external input voltage or current (±10 V voltage or 4 to 20
mA current) converted at the preset conversion factor.
EXT AVG2
*1
Shows the mean external input voltage or current (±10 V voltage or 4 to 20
mA current) converted at the preset conversion factor.
EXT INITIAL
*1
Shows the measured external input voltage or current (±10 V voltage or 4 to
20 mA current) before welding converted at the preset conversion factor.
Select CURRENT START or EXTERNAL INPUT for the timing setting
INITIAL MEASUREMENT in the EXTEND SETUP (7) screen.
EXT FINAL
*1
Shows the measured external input voltage or current (±10 V voltage or 4 to
20 mA current) after the end of current flow converted at the preset
conversion factor. Select CURRENT START or EXTERNAL INPUT for the
timing setting INITIAL MEASUREMENT in the EXTEND SETUP (7) screen.
EXT REAL TIME
*1
The MM-400A constantly measures the external input (±10 V voltage or 4
to 20 mA current) while MEAS is selected. The MM-400A stops
measurement when PROG is selected. Measurement is made at intervals of
twice a second. If you wish to select EXT REAL TIME for measurement,
select REAL TIME for TRIGGER in the BASIC SETUP (1) screen.
EXT TIME
*1
Shows the time from the start to the end of external input (±10 V voltage or 4
to 20 mA current).