MM-400A
8. Operation Screens
8-19
When 1x sensitivity coil is used:
2.000 kA range: 0.000 to 9.999 kA
6.00 kA range: 00.00 to 99.99 kA
20.00 kA range: 00.00 to 99.99 kA
60.0 kA range: 000.0 to 999.9 kA
200.0 kA range: 000.0 to 999.9 kA
When 10x sensitivity coil is used:
0.200 kA range: 0.000 to 9.999 kA
0.600 kA range: 0.000 to 9.999 kA
2.000 kA range: 0.000 to 9.999 kA
6.00 kA range: 00.00 to 99.99 kA
20.00 kA range: 00.00 to 99.99 kA
VOLTAGE (PEAK, RMS, AVERAGE RMS)
Set upper and lower limits in the following ranges according to the
VOLTAGE RANGE setting in the BASIC SETUP (1) screen:
6.00 V range: 0.00 to 9.99 V
20.0 V range: 0.0 to 99.9 V
WELD TIME
Set upper and lower limits in the following ranges according to the TIME
setting in the BASIC SETUP (1) screen:
CYC-AC, CYC***Hz-AC, CYC-DC, LONG CYC-AC: 0.0 to 3000.0 CYC
ms-DC, ms-AC: 0 to 30000 ms
SHORT ms-DC: 0.00 to 300.00 ms
POWER AVERAGE
0.0 to 300.00 kW
RESISTANCE AVERAGE
0.0 to 300.00 m
FORCE (AVERAGE 1, AVERAGE 2, PEAK, INITIAL, FINAL)
*1
Set upper and lower limits in the following ranges according to the SENSOR
and UNITS settings in the EXTEND SETUP (5) screen:
MA-520: 0.00 to 99.99 N / 0.00 to 99.99 kgf / 0.00 to 99.99 lbf
MA-521: 0.0 to 999.9 N / 0.0 to 999.9 kgf / 0.0 to 999.9 lbf
MA-522, MA-770A, MA-771A: 0 to 9999 N / 0 to 9999 kgf / 0 to 9999 lbf
When RATED SETTING1 or RATED SETTING2 is selected for SENSOR, set
upper and lower limits in the following ranges according to the DECIMAL
RANGE setting in the EXTEND SETUP (5) screen:
**.**: 0.00 to 99.99 N / 0.00 to 99.99 kgf / 0.00 to 99.99 lbf
***.*: 0.0 to 999.9 N / 0.0 to 999.9 kgf / 0.0 to 999.9 lbf
****: 0 to 9999 N / 0 to 9999 kgf / 0 to 9999 lbf
EXTERNAL (AVERAGE 1, AVERAGE 2, PEAK, INITIAL, FINAL)
*1
Set upper and lower limits in the following ranges according to the DECIMAL
RANGE setting in the EXTEND SETUP (7) screen:
*.***: -9.999 to +9.999
**.**: -99.99 to +99.99
***.*: -999.9 to +999.9
****: -9999 to +9999
DISPLACEMENT (INITIAL, PULSE 1, PULSE 2, FINAL)
*1
Set upper and lower limits in the following ranges according to the SENSOR
STEP and UNITS settings in the EXTEND SETUP (3) screen:
When a sensor with 1 μm resolution or less is used: -30.000 to +30.000 mm /
-3.000 to +3.000 inch
When a sensor with 1.1 μm resolution or more is used: -300.00 to +300.00 mm
/ -30.000 to +30.000 inch