Alarm Condition / Description / Corrective Action System
Alarm
Reference Gas Alarm Yes
Delta frequency is less than -0.3 Hz.
Corrective Action:
Verify analyzer. Check and/or replace dryer.
Contact AMETEK Service if problem persists.
Enclosure Temperature Yes
Electronics compartment temperature Exceeds 70 °C.
Corrective Action:
Contact AMETEK Service.
Moisture Generator Date Alarm No
Moisture generator is more than two (2) years old.
Corrective Action:
This indicator is given as a warning;
No user action is required.
Dryer Alarm No
Dryer capacity exceeded.
Corrective Action:
Replace dryer.
Concentration Alarm No
Moisture concentration exceeds the user-dened limits.
Corrective Action:
This is a user-dened alarm.
Analyzer Oine No
Active during verication cycle, and when any system alarm
is present.
Corrective Action:
Readings obtained when the invalid signal is active are not
a reliable measure of the Sample gas.