Chapter 7 Remote Commands
7-42
No.11 can be executed:
> :SENSe:MEASure:ASEarch:SLOT? USB11
< 1
Compatibility Incompatible with existing models.
:CALCulate:DATA:ASEarch? <slot>,<interface>[,<unit>]
Parameter <slot>=<CHARACTER PROGRAM DATA>
SLOT1 to SLOT6 Slots No. 1 to No. 6
1 to 6 when using the MP1800A, 1 to 4 when using the MT1810A
USB11 to USB14 USB No.11 to 14
<interface>=<DECIMAL NUMERIC PROGRAM DATA>
1 to 4 Data1 to 4
[<unit>]=<DECIMAL NUMERIC PROGRAM DATA>
1 to 4 Mainframe Nos. 1 to 4
When using two or more MT1810A units in serial connection, specify the
mainframe number.
Can be omitted. Mainframe No. 1 is specified when omitted.
Response <data>,<xdata>,<delay>,<numeric1>,<numeric2>,<numeric3>,<numeric4>
<data>,<xdata>,<delay>,=<STRING RESPONSE DATA>
"Done" Normal termination
"Fail" Abnormal termination
"----" Not executed
<numeric1>,<numeric2>=<NR2 NUMERIC RESPONSE DATA>
–3.500 to 3.300 –3.500 to 3.300 V/Step 0.001 V
"----" Not executed
"----" Abnormal termination
<numeric3>=<NR1 NUMERIC RESPONSE DATA>
–1000 to 1000 –1000 to 1000 mUI/Step 1 mUI
"----" Not executed
"----" Abnormal termination
<numeric4>=<NR2 NUMERIC RESPONSE DATA>
–xxx.xx to xxx.xx ps (This value is converted from mUI units.)
"----" Not executed
"----" Abnormal termination
Function Queries the Auto Search results.
Example To query the Auto Search results of USB No.11:
> :CALCulate:DATA:ASEarch? USB11,1
< "Done", "Done", "Done",1.000,1.100,–500,50.00
< "Done", "Done", "---",1.000,1.100,0,0.00 (When not executed in
the Phase direction)
Compatibility Incompatible with existing models. (The parameters are different.)