1.3 Specifications
1-7
1
Overview
Table 1.3.1-1 Specifications for MP1862A (Cont’d)
Item Specifications Remarks
Data Input Number of Input 2 (Data/
Data
Input) (Differential)
Input Conditions Selectable from Single-ended 50
Differential
50
, or Differential 100
Data or XData can be set to input for
Single-ended interface.
Tracking, Independent, or Alternate can be set
for Differential Interface.
When setting to Alternate, Data-XData or
XData-Data can be set.(Only when absolute
difference between Data Threshold and XData
Threshold is 3.0 V or less)
Input Format NRZ
Amplitude 0.125 to 1.0 Vp-p*
1,
*
5
Threshold –3.5 to +3.3 V /1 mV step
•
Data and XData can be set separately.
•
Absolute difference between Data Threshold
and XData Threshold : 3.0 V or less
Input Sensitivity
*
2,
*
3,
*
4,
*
6
Typ.25mV, ≤40mV@56.2Gbit/s
Typ.30mV@64.2Gbit/s
*
8
Phase Margin Typ. 200 deg*
7,
*
8
Termination 50
/GND, Variable
Termination
Voltage
When Variable is selected for Termination:
–2.5 to +3.5 V /0.01V step
Connector V (f.)
*1: Defined based on the condition where Single-ended 50
is selected.
*2: Defined under the following test pattern conditions:
• PRBS2^31–1
• Mark Ratio1/2
*3: Defined in the temperature range of 20 to 30°C.
*4: Defined by looping back a signal through J1600A coaxial cable,
41V-6 fixed attenuator, and MP1861A.
*5: The input amplitude corresponds to the operating range of the
Auto Search and Auto Measurement functions, and the lower limit
for the input amplitude is defined by Eye Height.
*6: This is determined by Eye Height.
The output amplitude must be that of MP1861A + ATT with an
error rate being “1E–9” or lower in the following measuring system
(a sampling oscilloscope with more than 70 GHz bandwidth is used
for output amplitude observations).