4.4 Setting Input Interface
4-13
4
CAUTION
Be sure to connect the supplied protection cap (J1363A) to
the data input connector you don’t use when the data
input condition is set to Single-ended.
Operating while signal is input to the unused connector
can cause malfunction.
Table 4.4.1-2 Data input condition setting items
Data Termination setting items Description
100Ohm
When the input connector is open, the midpoint of the Data
terminator (50
Ω
) and the XData terminator (50
Ω
) is connected
to the ground through a high resistance for equipment safety.
Ω
Terminates into a setting voltage in the range of 50 W/–2.5 to
+3.5 V, in 10 mV steps.
CAUTION
Avoid feeding excessively high current to the internal
termination resistor of MP1862A. Failure to do so may
degrade the performance or cause a malfunction.
If a differential signal is input to each of the Data and
XData connectors when Single-Ended is clicked in the
Input Condition box, the threshold margin will be
doubled.
Table 4.4.1-3 Threshold Voltage settings
Setting items Description
XData Threshold
Set in range of –3.500 to + 3.300 V in 0.001-V steps.
If
or
is clicked in the
box, values for
and
are restricted so that the difference (absolute value)
of them can be within 3.000 V.
Set in range of –3.000 to +3.000 in 0.001-V steps.