Chapter 3 Measurement
3-30
RF Power Set Reference
Summary
Adds the display of relative value results as the RF Power
measurement results. The RF Power measurement result at the
point when this setting is turned On is assumed to be the reference
(0 dB), and the relative values of the subsequent RF Power
measurement results are displayed.
Note:
This can be set to On only when AF Level measurement results are
displayed.
Options
On Displays the relative values of the RF Power
measurement results.
Off Does not display the relative values of the RF
Power measurement results.
DCS Analysis
Summary
Demodulates DCS (Digital Code Squelch) signal and displays the
DCS Code. This setting is enabled when Modulation is set to FM.
Options
Off Disables DCS Code analysis.
Normal Enables DCS Code analysis and displays the
analysis result.
Inverted Enables DCS Code analysis and displays the
analysis result as invert signal.
Limit Level
Summary
Sets the lower limit of RF Power to display the AF Measurement
result. If the RF Power signal is lower than the value set with Limit
Level, the AF Measurement result is not displayed.
Setting range
When Pre-Amp is Off:
(–60.0 + Input Offset Value) to (30.0 + Input Offset Value) dB
When Pre-Amp is On:
(–80.0 + Input Offset Value) to (10.0 + Input Offset Value) dB