3.4 TX Measurement Mode
3-49
3
Measurement
20 kHz Sets the minimum value of the X-axis scale to 20 kHz.
30 kHz Sets the minimum value of the X-axis scale to 30 kHz.
Stop Freq
Summary
Sets the maximum frequency for X-axis on the Frequency Domain
graph.
Setting range (X-AXIS is Linear)
60 to 50000 Hz
Options (X-AXIS is Log)
20 Hz Sets the maximum value of the X-axis scale to 20 Hz.
30 Hz Sets the maximum value of the X-axis scale to 30 Hz.
50 Hz Sets the maximum value of the X-axis scale to 50 Hz.
100 Hz Sets the maximum value of the X-axis scale to 100 Hz.
200 Hz Sets the maximum value of the X-axis scale to 200 Hz.
300 Hz Sets the maximum value of the X-axis scale to 300 Hz.
500 Hz Sets the maximum value of the X-axis scale to 500 Hz.
1 kHz Sets the maximum value of the X-axis scale to 1 kHz.
2 kHz Sets the maximum value of the X-axis scale to 2 kHz.
3 kHz Sets the maximum value of the X-axis scale to 3 kHz.
5 kHz Sets the maximum value of the X-axis scale to 5 kHz.
10 kHz Sets the maximum value of the X-axis scale to 10 kHz.
20 kHz Sets the maximum value of the X-axis scale to 20 kHz.
30 kHz Sets the maximum value of the X-axis scale to 30 kHz.
50 kHz Sets the maximum value of the X-axis scale to 50 kHz.
Top Level
Summary
Sets the maximum value of the Y-axis on the Frequency Domain
graph.
Setting range
For FM modulation:
(Setting value of Bottom Level
1.4125375) to 1000 kHz
For AM modulation:
(Setting value of Bottom Level
1.4125375) to 1000%
For
M modulation:
(Setting value of Bottom Level
1.4125375) to 1000 rad
Note:
The setting range of Top Level should meet the following
conditions.
Top Level Bottom Level
1.4125375
Top Level – Bottom Level 3 dB