3.4 TX Measurement Mode
3-51
3
Measurement
3.4.8.3 Setting Display Mode
This section describes the settings of the items on the Result window and
AF Measurement Result (TX-AF) window. Press
(Display Mode) on
the Trace function menu to display the Display Mode function menu.
ALL
■
Summary
Displays all measurement results.
Carrier
■
Summary
Displays the Result window only.
The Result window displays RF Frequency, Freq. Error, and RF Power.
Modulation
■
Summary
Displays the Result window and AF Measurement Result (TX-AF) window.
The Result window displays the following.
● Deviation (Modulation is FM or Wide FM )
● Depth (Modulation is AM)
● Radian (Modulation is
M)
The AF Measurement Result (TX-AF) window displays the following.
● Level
● Distortion or SINAD or THD.
3.4.8.4 Setting Carrier Display Mode
This section describes the settings for Carrier Display Mode. When
Display Mode is Carrier, press
(Carrier Setup) on the Trace function
menu to display the Carrier Setup function menu.
RF Power Unit
■
Summary
Sets the unit of RF Power measurement results.
■
Options
W Displays the RF Power result in W.
mW Displays the RF Power result in mW.
dBm Displays the RF Power result in dBm.
RF Frequency (Hz) Resolution
■
Summary
Sets the resolution of RF Frequency measurement result and Freq.
Error measurement result.