Usage specifications
7990044_030_10 – 2078511 – 2023-01-19 11
1 Usage specifications
1.1 Intended purpose
The operating table is intended for the following use:
Patient positioning during surgical interventions including
initiation and ending of anesthesia
Application-related transfer of patients within operating rooms
For use in combination with intra-surgical imaging (e.g. CT,
MRT, X-ray) with special table components
1.2 Contraindication
Do not transport objects, devices, or materials on the operating
table and its accessories. The products described in these
instructions for use are for human medical use only.
Hazard to the patient. Changes to the medical product are
prohibited. The manufacturer is not liable for changes made to the
operating tabletop.
Risk of personal injury and damage to goods when the permitted
loads are exceeded. Do not exceed the permitted load capacity
for the operating table. If the permissible load is exceeded, the
mobile operating table may tip over and cause serious injuries to
the patient or staff. In general, overloading the operating table can
lead to a failure of electrical functions and cause material damage
to mechanical parts.
Hazard to the patient. The patient must only be positioned on the
operating tabletop while lying down. Extremities must not extend
beyond the end of the operating table in a longitudinal direction.
Improper loading may damage the operating tabletop or cause
the mobile operating table to tip over. When the operating
tabletop is moved in a longitudinal direction, the patient must not
climb onto or off the operating table over the side of the extended
operating tabletop. Patients may only get on or off in the area of
the support column. Do not sit, crouch or kneel on the end of the
tabletop.
1.3 Patient definition
All patients who are classified as suitable for the operation by a
medical expert.
The maximum possible patient weight on the operating table is
450 kg / 992 lbs.
The maximum operating table weight of 300 kg / 661 lbs consists
of the operating table weight and the weight of the attached
tabletop sections.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter TruSystem 7000 and is the answer not in the manual?

baxter TruSystem 7000 Specifications

General IconGeneral
Brandbaxter
ModelTruSystem 7000
CategoryMedical Equipment
LanguageEnglish

Related product manuals