7990044_030_10 – 2078511 – 2023-01-19
movement to bring the operating table with its current load back
into a safe range. If the operating table load is exceeded, an error
message is shown on the remote control display. Only the flashing
evasive movement is permitted on the operating table.
Example: When checking the operating table load, the electronics
system detects a head-end overload of the operating table. The
warning or error message appears on the remote control display.
In order to get the operating table load back into balance, press
flashing key [i17]. When the warning message is shown, only the
function for longitudinal slide adjustment toward the head end is
locked. If the operating table load is exceeded, all other functions
of the operating table are locked with the error message.
Some typical operating tabletop configurations and the
associated load limits are listed below. Compliance with the
specifications prevents overloading of the operating tabletop.
Symbol explanation:
Notifications on the column keypad for the head position of the
patient.
3.11.1 Operating tabletop with head section, upper back section and universal section
1. An upper back section [A] and a head section [B] are attached
at the head end of the operating tabletop.
2. The universal section [C] is attached at the foot end of the
operating tabletop.
3. Normal patient orientation is active on the operating table.
3.11.2 Operating tabletop with head section, universal section and leg section
1. A leg section [A] is attached at the head end of the operating
tabletop.
2. The universal section [B] and a head section [C] are attached
at the foot end of the operating tabletop.
3. The patient orientation on the operating table is switched to
inverted.
C
A
B
Permissible load:
450 kg /
992 lbs
Patient weights of up to 450 kg / 992 lbs are
permitted when the operating tabletop's longitudinal
slide is in the level position.
250 kg /
551 lbs
Patient weights of up to 250 kg / 551 lbs are possible
without any restrictions applying to the operating
table functions.
Attention: Additional accessories on the operating table reduce
the permissible patient weight.
B
C
A
Permissible load on the operating tabletop:
200 kg /
440 lbs
Patient weights of up to 200 kg / 440 lbs are possible
without any restrictions applying to the operating
table functions. Attention: Additional accessories on
the operating table reduce the permissible patient
weight.