Safety
7990044_030_10 – 2078511 – 2023-01-19 17
Activating emergency mode:
The emergency mode has to be manually activated by the user:
1. Unlock the column keypad. Press key [i53] for this.
2. Press key [i20] on the column keypad for at least 10 seconds.
A warning tone sounds and constantly repeats. The
emergency mode is confirmed by a double tone and is then
activated.
The indicator [i20] on the column keypad lights up red
continuously and the indicator [i1] flashes green.
The operating table can be restored to its normal mode of
operation by switching it off and then on again.
2.7.2 Emergency unlocking of the parking brake
If the operating table cannot be unlocked with the operating units
after lowering the tabletop, leveling the tilt and centering the
longitudinal travel due to a defect, then it is possible to unlock it
manually.
1. Press the button for the emergency release [18] at the head
end of the running gear until the operating table is
completely jacked down. All jack props [26] are immediately
retracted.
2. You can now move the operating table.
3. Inform Baxter Technical Customer Service.
No functionality is lost if the operating table had no previous
defect and the emergency prop release was inadvertently
activated. However, resetting the props immediately is strongly
recommended.