7990044_030_10 – 2078511 – 2023-01-19 47
3.11.7 Operating tabletop with extension unit
1. The extension unit [A] is attached at the foot end of the
operating tabletop.
2. The orientation of the patient on the operating tabletop is
normal.
3.11.8 Operating tabletop with shoulder chair
1. Leg sections [A] can be attached at the head end of the
operating tabletop, as required.
2. The shoulder chair [B] with the head-positioning accessory
[C] is fastened to the foot end of the operating tabletop.
3. The orientation of the patient on the operating tabletop is
inverted.
3.11.9 Operating tabletop with pediatric extension section
1. No tabletop section is required at the head end of the
operating tabletop.
2. The pediatric extension section [A] is attached at the foot end
of the operating tabletop.
3. The patient orientation on the operating table is switched to
inverted.
A
Permissible load:
160 kg /
352 lbs
Patient weights of up to 160 kg / 352 lbs are possible
without any restrictions applying to the operating
table functions.
B
A
C
Permissible load:
225 kg /
496 lbs
A patient weight of up to 225 kg / 496 lbs is
permitted.
Note that additional accessory parts reduce the permissible
patient weight.
A
Permissible load:
40 kg /
88 lbs
A patient weight of up to 40 kg / 88 lbs is permitted.
Attention: Additional accessories on the operating table reduce
the permissible patient weight.