Technical data
7990044_030_10 – 2078511 – 2023-01-19 89
10.4 Electromagnetic compatibility
Notice: The operating table may not be used in direct proximity to any other devices. If this is required,
the operating table must be continually monitored to ensure its proper operation under these
conditions.
The operating table may only be operated with the cables supplied with the product. The use of cables
other than those specified may result in increased transmission or reduced interference immunity of the
TruSystem 7000 operating table.
The properties of this device, determined according to its emissions, allow for its use in the industrial
sector and in hospitals (CISPR 11, Class A). When used in domestic situations (for which Class B is normally
required according to CISPR 11), this device may not provide adequate protection from radio services.
Table 1 according to IEC 60601-1-2:2014
Guidelines and manufacturer's declaration – electromagnetic interference
The TruSystem 7000 operating table is intended for use in the electromagnetic environments as
specified below. The customer or user of the aforementioned operating table must ensure that it is
operated in an environments as described.
Measurement of interference
emissions
Compliance Electromagnetic environment –
guidelines
HF emissions in accordance with
CISPR 11
Group 1 The TruSystem 7000 operating table uses
exclusively HF energy for its internal
FUNCTIONS. Its HF emission is therefore
very low and interference with adjacent
devices is unlikely to happen.
HF emissions in accordance with
CISPR 11
Class A The TruSystem 7000 operating table is
suitable for use in establishments other
than domestic and those connected
directly to the PUBLIC LOW-VOLTAGE
NETWORK that supplies buildings used
for domestic purposes.
Harmonic Emissions as per
IEC 61000-3-2
Class A
Voltage fluctuation/flicker emissions
in accordance with IEC 61000-3-3
fulfilled
Table 2 according to IEC 60601-1-2:2014
Guidelines and manufacturer's declaration – electromagnetic immunity
The TruSystem 7000 operating table is intended for use in the electromagnetic environments as
specified below. The customer or user of the aforementioned operating table must ensure that it is
operated in an environments as described.
Immunity testing EN/IEC 60601-1-2
testing level
Compliance level Electromagnetic environment
– guidelines
Static electricity
discharge (ESD)
according to EN/
IEC 61000-4-2
±8 kV contact
discharge
±2 kV, ±4 kV, ±8 kV,
±15 kV air discharge
±8 kV contact
discharge
±2 kV, ±4 kV, ±8 kV,
±15 kV air discharge
Flooring should be made of
wood or concrete or have
ceramic tiles. If a floor is
covered with synthetic
material, the relative humidity
must be at least 30%.
Electrical fast
transient
disturbances/bursts
in accordance with
EN/IEC 61000-4-4
±2 kV for mains power
cables
100 kHz repeat
frequency
±2 kV for mains power
cables
100 kHz repeat
frequency
Mains power quality should
correspond to a typical
commercial or hospital
environment.