Technical data
90
7990044_030_10 – 2078511 – 2023-01-19
Attention: Distance for portable RF communications equipment and its peripherals
Do not use portable RF communications equipment (including peripherals such as an antenna cable and
external antennas) at a distance of less than 30 cm (12 in) from the TruSystem 7000 operating table,
including its cables that have been specified by the manufacturer. Otherwise, the functionality of the
system may be impaired.
Impulse voltages
(surges) in
accordance with
EN/IEC 61000-4-5
±0,5 kV, ±1 kV line to
line
±0.5 kV, ±1 kV, ±2 kV
line to earth
±0,5 kV, ±1 kV line to
line
±0.5 kV, ±1 kV, ±2 kV
line to earth
Mains power quality should
correspond to a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines pursuant to EN/
IEC 61000-4-11
0 % U
T
; 0.5 cycle
a)
0 % U
T
; 0.5 cycle
a)
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the TruSystem 7000
operating table requires
continued operation, even
during interruptions of the
power supply, then it is
recommended to connect the
TruSystem 7000 operating
table to an uninterrupted
power supply or a battery.
0 % U
T
; 1 cycle 0 % U
T
; 1 cycle
70% U
T
; 25/30 cycles
b)
70% U
T
; 25/30 cycles
b)
0 % U
T
;
250/300 cycles
b)
0 % U
T
;
250/300 cycles
b)
Magnetic field at
supply frequency
(50/60 Hz) in
accordance with
EN/IEC 61000-4-8
30 A/m 30 A/m Magnetic fields for the network
frequency should comply with
values commonly found in
commercial and hospital
environments.
a) at 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
b) at 0°
Comment U
T
is the AC mains voltage prior to applying the test level.
Table 2 according to IEC 60601-1-2:2014
Guidelines and manufacturer's declaration – electromagnetic immunity
The TruSystem 7000 operating table is intended for use in the electromagnetic environments as
specified below. The customer or user of the aforementioned operating table must ensure that it is
operated in an environments as described.
Immunity testing EN/IEC 60601-1-2
testing level
Compliance level Electromagnetic environment
– guidelines