Description
50
7990044_030_10 – 2078511 – 2023-01-19
3.13 Use in imaging system
The tabletop is fully radiolucent between the struts. Avoid
wrinkling the pad and other materials lying on it (towels, underlays)
in order to keep image artifacts to a minimum.
The cladding protection is not radiolucent.
X-ray area of the operating tabletop:
[A] X-ray area, foot end
[B] X-ray area, head end
[C] Column area (not radiolucent)
The X-ray region [A]/[B] on the operating tabletop is the region
between the struts of the end of the tabletop section and the
cladding of the operating table column. The X-ray region is
dependent on the longitudinal travel of the operating tabletop.
The column area [C] is not radiolucent.
Cassette rails for X-ray cassette:
There are rails on the inside of the back section arms to hold an
X-ray cassette (size 35.56 cm: (14 inch)). The X-ray cassette can be
pushed beneath the seat/back section. Push the X-ray cassette
into the fixing points on the inside of the struts. The cassette is not
fixed and may slide if the operating tabletop is not horizontal or is
moved. Please note this in order to avoid damage to the X-ray
cassette.
3.14 Driving modes
The operating table has a running gear. The driving behavior of the
operating table is regulated via four freely swiveling wheels, four
props and one additional 5th wheel in the middle under the
running gear. The following running gear functions are selected or
supported using the operating table operating units (remote,
column keypad):
Unlock the operating table (Unlock)
Freewheel (Unlock)
Directional travel (Assist)
Rotate operating table (Assist)
Directional travel with drive unit (only for MB/MBW operating
table version) (drive)
Operating table lock (Lock, jacks extended)
3.15 Internal power supply (battery operation)
Two lithium ion batteries provide the internal power supply. With
the set of batteries charged, use of the electric operating table
functions is guaranteed for 16 hours (total duration of movement
for electrical functions). The electrical functions on the operating
table are blocked when the rechargeable battery set is depleted
or almost completely depleted. In this case, the external power
supply must be connected.
A
B
C

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter TruSystem 7000 and is the answer not in the manual?

baxter TruSystem 7000 Specifications

General IconGeneral
Brandbaxter
ModelTruSystem 7000
CategoryMedical Equipment
LanguageEnglish

Related product manuals