Description
36
7990044_030_10 – 2078511 – 2023-01-19
3.9 Sounds
Various audible signals sound in conjunction with specific
operational procedures or statuses of the operating table.
In addition to the acoustic signals, a corresponding indicator is
shown in the display when using the TruSystem 7000 remote
control device.
External power supply [a3]
Illuminated green Power cable is connected to
the operating table.
External power supply available
(power supply connection).
–
Malfunction [a4]
Illuminated in red An error has occurred with the
operating table.
Operating table can be used in
a limited manner.
Note the indicator in the
remote control display.
Inform the Technical
Customer Service.
*1 Only for operating table version MB/MBW
Display Status Color Meaning Action
Action Description of audible signal
Switching on the operating
table
Ascending tone sequence
Switching off the operating
table
Descending tone sequence
Maximum overload limit, end
position or level position of
selected adjustment range
reached
Single tone
Confirmation of an operational
procedure, e.g.:
– Key lock released Double tone
– Level position reached Double tone
– Running gear procedures
(travel, jacking up/down)
Repeating single tone
During movement the single
tone repeats at intervals of a
few seconds.
Operating table needs charging 2 pulsing tone sequences
intermittently repeating (at an
interval of several minutes -
battery tone)
Error Shrill triple tone (error tone)
Warning, e.g.:
– Leg sections moving toward
one another during single
hinge adjustment
–Emergency mode
Repeating high-pitched single
tone (warning tone)
During movement the warning
tone sounds repeatedly at
intervals of a few seconds.