Description
30
7990044_030_10 – 2078511 – 2023-01-19
The hook coupling of each tabletop section is described in the
instructions for use for each tabletop section.
The available hook couplings on the operating tabletop are
specified in Chapter 3.1.
3.5 Side rails
The operating tabletop is equipped with side rails. Accessories can
be attached to the side rails. The accessories approved by Baxter
are listed in the Compatibility Matrix
2)
. The accessories must be
taken into account when determining the total patient weight.
The side rails are available in various country-specific standards
and differ in their dimensions. The side rails have a safety
mechanism at the end [A]/[B]. The securing device prevents loose
accessories from sliding off the side rail.
3.6 Connections for cables
Only the control units and cables listed may be connected to the
operating table.
Connection socket [12] under the seat section:
– Cable remote control
–Footswitch
Connector pin [14] on the running gear:
– Equipotential bonding cable [20]
The operating table may be used only with the original
equipotential bonding cable from Baxter.
Connection socket [15] on the running gear
– Mains power cable [19]
The operating table must be connected to the power supply
using the original mains power cable from Baxter. The mains
power cable is marked with a label.
2 ) Document number 7990090