Use
7990044_030_10 – 2078511 – 2023-01-19 53
4Use
4.1 Safety instructions
Commissioning:
Before using the products for the first time, they must first be
cleaned and disinfected according to the hygiene
specifications of the medical facility. Cleaning and disinfecting
may be performed only by trained staff and using cleaning and
disinfecting agents approved by Baxter.
The medical device is subject to special precautionary
measures regarding electromagnetic compatibility (EMC).
Installation and commissioning must be in accordance with the
Chapter 10.3 and Chapter 10.4.
Safety during operation:
Follow the relevant instructions for use for proper and safe use
of additional equipment.
Risk of material damage to electrical equipment due to
condensation! There is a risk of condensation forming during
transport, which could potentially damage electronic
components. After delivery of the operating table and before it
is switched on or connected to an external power source, it is
important to wait a certain period of time (at least 12 hours is
recommended). Place the operating table in an environment
with the same ambient temperature and relative humidity as
the room where the table will be used. The batteries must be
completely recharged after a significant change in
temperature or humidity.
The manufacturer will not be held liable if installed protective
devices are removed, or if additional tabletop sections or
accessories that do not meet the manufacturer's specifications
are installed.
Check that all electrical and mechanical functions/parts of the
operating table, including accessories, are undamaged and in
good working order before use. The use of faulty or damaged
products is prohibited.
Hazard to the patient due to improper attachment. Check that
the section segments and accessories are securely fastened to
the operating table before each use. Spontaneous movement
may occur and loose elements may slip out. Loosen controls for
adjustment purposes only. Tighten them again immediately
after the adjustment. Check if fastening elements are properly
seated.
Risk of crushing by moving parts. Familiarize yourself with its
operating functions before use.
Risk of injury due to protruding parts. Keep sufficient distance.
Use of the operating table is permitted in combination with
other products from Baxter. The instructions for use of the
products used must be observed and followed when the
products are used.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter TruSystem 7000 and is the answer not in the manual?

baxter TruSystem 7000 Specifications

General IconGeneral
Brandbaxter
ModelTruSystem 7000
CategoryMedical Equipment
LanguageEnglish

Related product manuals