Description
52
7990044_030_10 – 2078511 – 2023-01-19
3.19 Software settings
The following settings can be configured on the operating table
via an Ethernet connection from a PC to the operating table by
using the Baxter Service Interface software:
1. Stop on passing through the level position (activated for all
functions by default):
All operating table electrical adjustment functions stop
automatically when the level position is reached. To adjust
beyond the level position, briefly release the function key and
press it again.
2. Restricted leveling (not active by default):
This function allows the user to adjust the Level position
function. In particular, the automatic leveling of the
motorized tabletop section joints can be prevented.
3. Interim stop for table tilt, Trendelenburg and longitudinal
travel (not activated by default):
In addition to the general level position, interim stops can be
set for the Tilting, Trendelenburg and Longitudinal travel
functions. The respective function stops automatically upon
reaching the newly set position. For any further adjustment
briefly release the function key and then press it again. For
safety reasons, the intermediate stop of the tilt function
cannot be deactivated (the intermediate stop can be at an
inclination angle from 5° to 15°).
4. Signal tone volume:
This function allows the volume of the signal tones on the
operating table to be individually adjusted.
5. Speed, drive unit:
This function enables individual adjustment of the speed of
the drive unit for the operating table version MB/MBW.
The desired option has to be adjusted on site by the Baxter
Technical Customer Service or the responsible hospital
technician.