BD BACTEC™ FX40 Instrument User’s Manual
146
47
Instrument X:
Agitation Restarted
Agitation speed is outside of
normal range, or failed to stop
at a sensor or see a sensor
for 10 continuous seconds.
A Drawer Open,
Measurement scan or Power
failure resets 4 times
consecutively.
Alert is reported on System Alerts
display and report only.
Alert is cleared when agitation speed
returns within range.
If this message recurs frequently,
contact BD for service.
49
FX40 ‘X’ Rows A &
B are Offline.
Remove any vials
in unusable
stations. Consult
Manual.
When the BD BACTEC™
FX40 Control Board fails to
communicate with the Row
Board that controls rows A
and B, the Row Board is
marked offline.
Reboot instrument.
Stations are marked as Unusable.
Refer to Section 7.3.2 for additional
information on dealing with Unusable
stations.
Move vials in the indicated stations
within 40 minutes of the set time of this
alert to prevent them from becoming
Affected Vials.
50
FX40 ‘X’ Rows C &
D are Offline.
Remove any vials
in unusable
stations. Consult
Manual.
When the BD BACTEC™
FX40 Control Board fails to
communicate with the Row
Board that controls rows C
and D, the Row Board is
marked offline.
Reboot instrument.
Stations are marked as Unusable.
Refer to Section 7.3.2 for additional
information on dealing with Unusable
stations.
Move vials in the indicated stations
within 40 minutes of the set time of this
alert to prevent them from becoming
Affected Vials.
Error
No.
Message Possible Cause(s) Corrective Action(s)