7 – Troubleshooting
159
WE34
FX40 selected is
currently offline.
Information shown
may not be
up-to-date.
Appears if an instrument that
is offline is selected in View
Stations or Block/Unblock
Stations display.
The instrument, a station, or a
row board could be what is
offline.
Message is informational. Vial or
station statuses may be different from
what is shown on the display because
the instrument cannot communicate
with the offline instrument.
WE35
Report is
estimated to be
more than 50
pages. Print
Report?
The instrument has calculated
the approximate size of the
report to be more than 50
pages.
To accept the message, select Yes.
Select No to cancel the print request.
WE36
FX40 contains
anonymous vials
that are in
unusable stations.
Print Maintenance
QC Report or
access the View
Stations Screen to
locate vials, then
identify using ID
Anonymous
workflow. Consult
Manual.
An instrument that contains
anonymous vials in unusable
stations was opened.
Use the ID Anonymous activity to
identify any vials in the instrument.
When identifying anonymous vials in
this scenario, be sure to either select
Save after identification to move them
to another instrument, or place the
vials in another station in the
instrument that is lit steady green.
WE53
FX40 contains one
or more vials that
are partially
seated. Fully insert
Vials?
The instrument has
determined that a vial may be
partially seated in the station.
Message is displayed when
the door is first opened and
each time it is subsequently
opened, until a measurement
occurs that clears the partial
insertion condition.
Refer to Section 7.3 for additional
information.
WE56
One or more FX40
doors is ajar.
Please close or
fully open the
door(s).
The door sensor has detected
that the door is not fully
closed.
Push door fully closed.
WE57
Database
under-write –
displayed data not
current. Please
change data again
and reattempt
save.
While you were attempting to
enter a new vial, identify an
anonymous vial, or change
vial or specimen information,
another process changed
information for that vial or
specimen.
Your current modifications are not
saved.
On the Culture display, recall the vial/
accession and modify the desired
information again.