BD BACTEC™ FX40 Instrument User’s Manual
160
WE58
Vial can no longer
be found in the
database and
cannot be viewed
or modified.
Consult manual.
This will be displayed if the
user has attempted to modify
a vial that has been deleted
by BD EpiCenter™ since the
time it was recalled on the
Culture screen.
Message is informational. No activity is
possible since the vial is no longer
resident in the database.
WE59
Vial last known to
be in station
shown above,
which is offline. Is
this the same vial?
Sequence vial is removed
from a degraded mode
instrument or offline row and
inserted via Vial Entry into an
online instrument.
Select Yes to remove the vial from the
old (offline) location.
Place the vial into the online
instrument to continue testing the vial
in its current protocol.
Select No to re-enter the vial
sequence number (in case sequence
was entered incorrectly).
WE60
Vial currently
residing in an
offline instrument
and cannot be
modified. Consult
manual.
In a BD EpiCenter™
configuration, you tried to
modify a vial in the Culture
display (e.g., disassociate or
associate accession, change
protocol length, select media
type (if replacement vial),
change status) if the vial
resides in an instrument
which is currently offline.
You cannot modify information for a
vial that is in an offline instrument.
WE61
Vial removed from
station shown
above is
POSITIVE.
Message occurs when a
removed vial is called P. This
will likely be the result of an
anonymous vial being
identified and saved (i.e. not
returned to the instrument).
Algorithms are re-run on the
vial with the newly identified
media type, and if more
sensitive than the general
algorithm, could cause a
positive result.
Message is informational. Vial is
already removed from the instrument.
WE62
Restoring
connection to
database.
The instrument and
BD EpiCenter™ systems are
reconciling their databases.
Message is displayed while the
reconciliation process is in progress.
Message is removed when the
reconciliation process is complete.
Error
No.
Message Possible Cause(s) Corrective Action(s)