EasyManua.ls Logo

BD BACTEC FX40 - Page 32

Default Icon
174 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BD BACTEC™ FX40 Instrument User’s Manual
32
location, you should leave this value set at 1. Instrument numbers appear in report headers.
Instrument numbers must be unique in a BD EpiCenter™ configuration.
BD BACTEC™ FX40 Serial Numbers
Read-only fields showing the instrument Serial Numbers. This number is set at instrument
installation. A, B, C, and D represent up to 4 instruments that can be configured in a local
cluster. If there is only one instrument, A is used.
An icon appears to the right of each Serial Number. The icon is red (see Figure 2-6) if a device
is not connected to the Network port. The icon is green if a device is connected to the Network
port.
Figure 2-6 – Instrument Configuration Display (BD BACTEC™ FX40 Serial Number icon red)
Address
Read-only fields showing network configuration information.
ID shows the original MAC address of the network interface card
NIC shows the current MAC address of the network interface card
IP shows the Internet protocol address
Select the volume of the Positive Vial audible alarm. The default setting is 8. To increase or
decrease the volume, select
or to increase or decrease the value in the field from 1
(softest) to 10 (loudest).
Select Volume to hear a sample of the current volume.
The volume of other instrument audible alarms is controlled by the system software and
cannot be adjusted.
Volume

Table of Contents

Related product manuals