4-35
Method 2: The result of impedance check for each electrode is passed.
Method 3: The EMG value range is from 65 to 75, and SQI value is 100
displayed on [Graphic Trends]
RM test(you can select either method to perform the test)
Method 1: The RM airway pressure is 58.2~61.8 cmH
2
O.
Method 2: The difference is within ±3% of reading when -20, 0, 10, 60,
120cmH2O is set for RM airway pressure accuracy test.
CCO/SvO
2
test
The CCO/SvO
2
numerics displayed on the patient monitor and Vigilance
monitor are consistent.
The waves (at the ECG signal output port) displayed on the oscilloscope
are consistent with the ECG calibration waves displayed on the monitor
screen.
The amplitude of electrical level at the signal output port of MAP, CVP
and SpO
2
are 5±0.25 V, 5±0.25 V and 10±0.5 V respectively.
PiCCO test(you can select either method to perform the test)
The pArt value displayed on the monitor is within 200±4 mmHg, and the
pCVP value is within 20±1 mmHg.
The waveforms of pArt and pCVP are displayed correctly.
Method 1: The BT value is 37±0.1ºC.
The value of
CatheterType is PV2015L20
The C.O. value is displayed on the monitor
Method 2: The BT value is 37±0.1ºC.
The value of
CatheterType is PV2015L20
The TI value is from 0 to 30ºC
NMT test(you can select either method to perform the test)
Method 1: The voltage change is detected by the multimeter and the
output of NMT stimulation is normal
Method 2: The pulse width is range from 180 to 220us, and the
calculated Stimulation Current is range from 33 to 37mA
The sensor check is pass.
EEG self-made module test (you can select one of these methods to perform the test)
Method 1: Sine wave can be normally displayed.
Method 2: It passes the impedance detection.
Method 3: The impedance displayed on the monitor is 5±1kΩ.
ANI test (you can select one of these methods to perform the test)
Method 1: The ANI value measured on a healthy adult is within 12–100.
Method 3: The impedance displayed on the monitor is within 57–63.
Nurse call relay performance test
The relay contacts are close when an alarm occurs.
Analog output performance test
The waves displayed on the oscilloscope are identical with those
displayed on the monitor.
Electrical safety test
The Power Plug and cord check is passed.