VC-6000
TM
Compact monitor Part C Sensor block
- 10 - C100583.002 Components Vers. 07 © VC-6000
TM
Compact monitor Sept. 2011
<OK latching>
This parameter defines whether the OK LED will operate in a latching or
non-latching mode. The setting can be changed at the instrument.
Settings:
OK latching Yes No
<Hardware Id>
This parameter informs you about the type of the 1-channel input
module. The setting is determined by the application firmware of the
VIBROCONTROL 6000 Compact monitor and cannot be changed at the
instrument.
Hardware-Id
Module
1
A-TIM (CCS)
2
A-TIM (-24 V)
3
D-TIM
6
V-TIM (15 Hz)
7
V-TIM (8 Hz)
1.2.2 Summary of the parameters
Parameter name Value range changeable
Socket no. 1 to 3 No
Sensitivity -99999.99...+ 99999.99 Yes
Input range 0...+ 99999.99 Yes
OK high -21.5 V...+2.5 V Yes
OK low -21.5 V...+2.5 V Yes
OK latching Yes / No Yes
Hardware Id 1, 2, 3, 6 or 7 No

Table of Contents