84
HYPERION User Manual Bruker Optik GmbH
Operation 4
13. In case of a motorized stage:
13.1. Save the current x/y-stage posi-
tion.
13.2. Put the supplied gold mirror on
the stage and move the stage in x- and/
or y-direction until the mirror is in the
beam path. Do NOT change the posi-
tion of the sample on the stage!
13. In case of a manual stage:
Put the supplied gold mirror on the
stage and move the stage in x- and/or
y-direction until the mirror is in the
beam path.
Motorized stage:
Step 13.1. corresponding OPUS function
Step 13.2. Joystick / corresponding
OPUS function
See OPUS/VIDEO Manua.
Manual stage: Rotary knobs (A and B in
fig. 3.6)
14. Focus on a mirror edge or on a
scratch of the mirror surface.
Note: The mirror surface needs to be
clean! Do NOT change the knife-edge
aperture setting!
Manual stage and motorized x/y-
stage: coarse and fine focus knobs at
the microscope (B/B’ and C/C’ in fig. 3.7)
Motorized x/y/z-stage: Joystick and
corresponding stage control functions in
OPUS
See OPUS/VIDEO Manual.
15. Activate measuring mode. either using the IR button on the operat-
ing panel of the microscope (C in
fig. 3.4) or in the OPUS video wizard
See OPUS/VIDEO Manual.
16. Start background measurement. corresponding OPUS function
See OPUS/VIDEO Manual.
17. Activate viewing mode. either the VIS button on the operating
panel (B in fig. 3.4) or in the OPUS video
wizard
See OPUS/VIDEO Manual.
Operating step Corresponding operating element(s)

Table of Contents