94
Standard-based Zeta Quantification
This section describes how to perform standard-based quantification of electron transparent
samples using beam current and thickness of standard sample as reference values.
Click the Standards configurator’s
calibration icon
Measure and enter the beam current
value as system factor.
Follow section 5.10 for electron transparent
samples. Ente
r the beam current in [pA] units
Check or Load the appropriate Zeta
factors.
use the Spectrum chart icon to
Enter the beam current as spectrum
parameter:
highlighted line
of the active spectrum in the spectrum list
(
below the spectrum) to open the
the beam current in pA as system factor in
the Parameter tab.
c) drag and drop a spectrum from project
to Spectrum chart
Select a TEM quantification method.
Alternatively drag and drop a quantification
method from the project to spectrum.
Set the quantification method to
interactive ( ).
access the QUANTIFICATION dialog.
Set Quantification model to Zeta factor
method
and perform quantification of
See section 5.12.1 to optimize quantification
and to save or export quantification method
and results.
Calculated thickness is shown below the
result table. Density is calculated based on
element composition. If density is known, it
can be entered and sample thickness wi