User Manual 4 The ESPRIT 2 Software
26
4.2 User Administration
Fig.
4.2-1 QUANTAX communication server
access
Fig. 4.2-2 QUANTAX user list
Fig. 4.2-3 Assigning a new user
The QUANTAX communication server
controls the
user assignments as well as the internal and external
client-
server communication. Service tools and
information for troubleshooting are also
provided.
The communication server runs on the QUANTAX
server computer. It is also present in single computer
and standalone systems.
Clicking the Bruker icon in the Windows
®
system
tray of the server computer opens the console of the
communicati
on server. With exception of the actions
described below no changes to the communication
server settings must be made. Unauthorized altering
of settings can render the system useless and lead
to severe data loss.
New users
are to be assigned from the QUANTAX
communication server. Double clicking the Bruker
icon in the Windows
®
system tray opens the
console of the communication server.
The tab QUANTAX user
lists the current assigned
users. A new u
ser can be added after clicking the
button. A password can be entered (leaving
the field empty is permitted) and an optional user
description can be added. As user type User, Main
user, and Supervisor can be selected.
User has limited access to the ESPRIT System
workspace
. This user level only allows access to the
Appearance
tab and can review the connected
instruments and license.
The Main user
has limited access to the ESPRIT
System workspace.
This user level has full access to
the tabs System, Appearance
and can calibrate the
connected EDS detectors as well as the imaging
system. The Main user
has no access to the service
settings of the WDS detector and the
electron
microscope driver.
The Supervisor has full access to all tabs in
the
ESPRIT System workspace.
Each new user is automatically assigned a folder
structure in a private volume of the server. Default
methods, a predefined user profile, and demo data
are copied to the user folders.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bruker QUANTAX EDS and is the answer not in the manual?

Bruker QUANTAX EDS Specifications

General IconGeneral
BrandBruker
ModelQUANTAX EDS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals