User Manual 5 Step-by-Step Guides
88
5.12.4
Standard-based Quantification with Standards Library
Standard-based quantification using a standards library allows the measurement of standard and
unknown samples at different times using beam current dependent system factor.
Step
Example/hints
1
Click the
icon in the bottom left
corner of the
Standards configurator
to
open the
ADMINISTRATION OF
STANDARD LIBRARY dialog and
Follow section 5.11.1 to measure standard
samples and update
the standards library.
a) Load a standards library
b) Create a New standards library.
2
Perform system factor calibration.
Refer to section 5.10.
3
Select
the or
workspace and
This workflow also applies to any spectrum
chart available in the ESPRIT software.
a) a spectrum
Follow section 5.7.
b)
Use the Spectrum chart icon to
open a spectrum.
Optionally, drag and drop a spectrum from
the
Project into the spectrum workspace.
4
Click
on the
button and
select
Load… to load a quantification
method.
Select a standard-based quantification
method,
e.g.: Linemarker Standard.mtdx.
5
Click
.
Follow section 5.12.1. Make sure, that
Use
standards
is selected under
Quantification
model in the method editor panel.
6
Use the
button of the result table to
Co
py, Save or Add to Project or
Add to
report.
Save the modified quantification method for
further use. Use
on the
button or
Save in the quantification dialog.