Model 201 Installation & Technical
8400-M117-O1 Rev D
Page 131
AUtoP= (Automatic Print)
This setting controls whether the 201 Admix will automatically print on
all ports set to printer output upon reaching a stable target weight.
NOTE: If the DFC Mode is set to “Off” then this setting is not applicable.
Press the F3/ key to show the current setting. If the setting displayed is
acceptable, press the F3/ key again to save it. Otherwise, use the
F2/ key to toggle to a new setting and then press the F3/ key to save
it. Allowable settings are:
no
no (No Automatic Print)
YES
YES (Automatic Print after stable target weight)
Pulse Output Operation Information and Recommendations
Pulse output operation is enabled when DFC Mode is set to off and the
Pulse Target setting is greater than zero. A pulse will ONLY be generated
for every positive increment of the Pulse Target. Upon power up, the 201
Admix system will calculate an offset if the scale does not power up at
zero weight. This offset will be used to calculate Pulse Target increments
until the scale is zeroed. Once zeroed, the scale will reset the increments
so that they start from zero instead of the initial offset. For this reason, it
is recommended that the scale be zeroed before each filling operation
when the pulse output is used by a controller that is driving the I/O to
guarantee the most accurate count of pulses in a particular filling cycle.
At Zero Output Operation
The “at-zero” output will be activated when the scale goes below the zero
tolerance amount and will remain active. If At Zero Delay (0-dEL on
display) is set to something other than zero, then the indicator will pause
for the allotted time before activating the output. Refer to the fast=
(Fast Target) parameter.