Model 201 Installation & Technical
8400-M117-O1 Rev D
trL= (4% Zero Range)
Press the F3/ key to show the current setting. If the setting displayed is
acceptable, press the F3/ key again to save it. Otherwise, use the F2/
key to toggle to a new setting and then press the F3/ key to save it.
CALPt= (Number of Calibration Points)
Calibration of the weight transmitter can be performed using multiple
calibration points to obtain a more precise weight reading over the entire
span of the scale.
Press the F3/ key to show the current setting. If the setting displayed is
acceptable, press the F3/ key again to save it. Otherwise, use the
F2/ key to toggle to a new setting and then press the F3/ key to save
it. Allowable values are 1 to 4.
mNOTE: The CALPt= parameter is used to set the calibration
points in addition to the zero (no load) weight calibration point.
DO NOT include the zero (no load) weight calibration point when
selecting the number of calibration points. For example, selecting
2 for the CALPt= parameter, you would have a total of three
calibration points, the zero (no load) weight and two calibrations
points that could correspond to a mid-point weight and the test
(load) weight.