EasyManuals Logo

Cardinal 201 User Manual

Cardinal 201
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #85 background imageLoading...
Page #85 background image
Model 201 Installation & Technical
8400-M117-O1 Rev D
Page 77
D2A= (Minimum Weight 0 volts)
This is the value, in weight, which outputs 0 volts from the DAC. All
weight below this target will output 0 volts.
Press the F3/ key to show the current setting. To accept the setting
displayed, press the F3/ key again to save it. Otherwise, use the
F2/ and F1/ keys to enter a new setting and then press the F3/ key
to save it. Allowable values are: 0 through 999999.
NOTE: If a negative weight is required for the minimum weight,
you must use the Web Configuration Page.
D2b= (Maximum Weight 10 volts)
This is the value, in weight, which outputs the maximum output 10 volts.
All weights above this value will output maximum volts from the DAC.
Press the F3/ key to show the current setting. To accept the setting
displayed, press the F3/ key again to save it. Otherwise, use the
F2/ and F1/ keys to enter a new setting and then press the F3/ key
to save it.
This weight must be a positive value, up to capacity of scale, and above
the d2a= value. Allowable values: 0 through 999999.
AdC= (Analog Input)
Press the F3/ key to show the current setting. If the setting displayed is
acceptable, press the F3/ key again to save it. Otherwise, use the
F2/ key to toggle to a new setting and then press the F3/ key to save
it. Allowable settings are:
off
Off (Analog Input is disabled)
uoIt
Read Voltage
Currnt
Read Current

Table of Contents

Other manuals for Cardinal 201

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Cardinal 201 and is the answer not in the manual?

Cardinal 201 Specifications

General IconGeneral
BrandCardinal
Model201
CategoryTransmitter
LanguageEnglish