Loading...
8545-M838-O1 Rev D 825 Installation & Technical
13
INSTALLATION, CONT.
9. Referring to the table below and the labels on the circuit board for terminal connections,
connect each wire to the terminal block.
Load Cell Wiring Table
Board Label
Function
Board Label
Function
+EXC
+ EXCITATION
-SIG
- SIGNAL
+SEN
+ SENSE
-SEN
- SENSE
+SIG
+ SIGNAL
-EXC
- EXCITATION
SHLD
SHIELD (Not used, the shield wire is connected to metal gland).
10. To terminate a wire, loosen the screws in the terminal block and then insert the wire into
the terminal opening. Tighten the screw to secure the wire in place. See Figure No. 7.
11. Repeat the procedure until all wires are in place.
12. After all terminations have been made, remove the excess cable from the enclosure.
13. Referring to Figure No. 8, fold the shield wire back over the plastic insert and then insert
the plastic insert (with the shield wire) into the gland connector.
Figure No. 8
Shield
Wire
Insert wire and
tighten screw.
Figure No. 7

Table of Contents