8545-M838-O1 Rev D 825 Installation & Technical
SETUP AND CALIBRATION, CONT.
To Begin Setup and Calibration
1. Press the ON/OFF key to turn on the 825. The display will perform a short self-test and
then change to the Startup screen showing the software versions and status of the main
board and option cards.
NOTE: When the 825 is initially started, a countdown starting at 7 seconds will
appear to the right of the 1. Application program selection. When the
countdown is complete, the application program mode will automatically start.
2. Press the Navigation Keys Down Arrow or UP Arrow to stop the countdown to allow
unlimited time to review the Startup screen.
3. The screen will display the status of the main board and the option cards. The status will
show “good” if there is no error indicated or an error code if there is an error. Refer to the
Error Codes section of this manual for more information.
4. If no error code is shown, press the Navigation Keys Down Arrow or UP Arrow to
(highlight) the 3. Legal metrology information / Setup option and then press
the ENTER key or press the 3 key to select the 3. Legal metrology information /
Setup option.
5. The display will change to show the Legal metrology information / Setup
screen.