8545-M838-O1 Rev D ļ‚Ÿ 825 Installation & Technical
39
SETUP AND CALIBRATION, CONT.
Figure No. 22
NOTE: With the calibration jumper installed, the display will show:
Hardware calibration disabled
after selecting 3. Calibration on the Setup Menu, and 1. Scale 1 on the
Setup Scale Number screen.
If you continue to try to perform calibration by selecting 2. Calibrate on the
Scale 1 – Calibration Options screen, the display will show:
CALIBRATION SEAL
Calibration locked by hardware jumper
Inside enclosure.
Press any key
Jumper J10 OFF
Calibration Enabled
Jumper J10 ON
Calibration Disabled
CAL INHIBIT
Jumper J10