EasyManua.ls Logo

Cardinal 825 - Total Calibration General Settings; Total Base Units Configuration; Total Interval Settings

Cardinal 825
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8545-M838-O1 Rev D 825 Installation & Technical
63
SETUP AND CALIBRATION, CONT.
Total Calibration General
Base Units:
With the Total – Calibration – General screen displayed the current setting for the
Base Units: parameter will be shown. If the setting displayed is acceptable, proceed
to the next parameter.
Otherwise, press the SPACE key to select the Base Units and then press the Down
Arrow to advance to the next parameter or press the ENTER key to save the new setting
and proceed to the next screen.
Interval:
With the Total – Calibration – General screen displayed the current setting for the
Interval: parameter will be shown. If the setting displayed is acceptable, proceed to
the next parameter. NOTE: If using multiples intervals, this is the low interval.
Otherwise, use the numeric keys to enter the interval (or minimum increment of weight
reading) and then press the Down Arrow to advance to the next parameter or press
the ENTER key to save the new setting and proceed to the next screen. Allowable
values are 1, 2, 5, 10, 20, 50 or 100.
Interval Dec:
With the Total – Calibration – General screen displayed the current setting for the
Interval Dec: parameter will be shown. If the setting displayed is acceptable,
proceed to the next parameter.
Otherwise, use the numeric keys to enter the interval decimal places and then press the
Down Arrow to advance to the next parameter or press the ENTER key to save the
new setting and proceed to the next screen. Allowable values are 0 to 5.
0 = #######
2 = #####.##
4 = ###.####
1 = ######.#
3 = ####.###
5 = ##.#####
Conv Units:
With the Total – Calibration – General screen displayed the current setting for the
Conv Units: parameter will be shown. If the setting displayed is acceptable, proceed
to the next parameter.
Otherwise, press the SPACE key to select the Conv Units and then press the ENTER
key to save the new setting and proceed to the next screen.

Table of Contents

Related product manuals